Olympus OMNISCAN X3 Phased Array Flaw Detector
The OmniScan X3 flaw detector is a complete phased array toolbox. Powerful tools, like total focusing method (TFM) images and advanced visualization capabilities, enable you to complete your inspection with greater confidence.
Innovative TFM
– Better flaw imaging brings clarity and resolution to smaller defects
– Image early-stage high-temperature hydrogen attack (HTHA) to detect it when it matters most
– Onboard acoustic influence map (AIM) reflectivity simulator helps you visualize TFM sensitivity and adjust accordingly
– Up to four TFM modes facilitates flaw interpretation and sizing
Improved Phased Array
– Twice as fast as the OmniScan MX2 (pulse repetition frequency)
– Single TOFD menu for accelerated workflow
– Improved fast phased array calibration minimizes frustration
– 800% high amplitude range reduces the need to rescan
– Onboard Dual Linear Array and Dual Matrix Array probe support accelerates setup creation
Get to Work, Quickly
The onboard scan plan, improved fast calibration, and streamlined user interface help eliminate unnecessary steps so that you can complete your inspection setup in minimal time.If youre an existing OmniScan user, the transition from the MX2 is fast. If youre new to phased array ultrasonic testing or TFM, the OmniScan X3 flaw detector is easy to learn.
The Reliability You Trust
– IP65 certified rain and dust proof
– Onboard GPS provides the location of data
– Connect wirelessly to the Olympus Scientific Cloud to download the latest software
– Scan without stopping thanks to the 25 GB file size
Specifications
Type
Multigroup, multimode ultrasonic flaw detector |
Size (W H D) |
335 mm x 221 mm x 151 mm (13.2 in. x 8.7 in. x 5.9 in.) |
Weight |
5.7 kg (12.6 lb) (with 1 battery) |
Hard Drive Capacity |
Internal 64 GB SSD, extendable using an external USB drive |
Storage Devices |
SDHC and SDXC cards or most standard USB storage devices |
Max Onboard File Size |
25 GB |
GPS |
Yes (unless specified otherwise for some regions) |
Alarms |
3 |
Wireless Connection |
Yes (USB dongle sold separately as an accessory) |
PA Connectors |
1 connector |
UT Connectors |
4 (2 channels P/R) for 32:128PR and 16:128PR, the 16:64PR has the same connectors but only one UT channel active (P1/R1) |
Certifications |
ISO 18563-1:2015 EN12668-1:2010 |
Display
Type |
TFT LCD with resistive touch screen |
Size |
269 mm (10.6 in.) |
Resolution |
1280 768 pixels |
Number of Colors |
16 million |
Viewing Angles |
Horizontal: 85 to 85 Vertical: 85 to 85 |
I/O Ports
USB 2.0 |
2 ports (one hidden behind the battery) |
USB 3.0 |
1 port |
Video Output |
Video out (HDMI) |
Memory Card |
SDHC port |
Communication |
Ethernet |
I/O Lines
Encoder |
2-axis encoder line (quadrature or clock/direction), 3rd encoder ready |
Digital Input |
6 digital inputs, TTL |
Digital Output |
5 digital inputs, TTL |
Acquisition On/Off Switch |
Through the configuration of a digital input |
Power Output Line |
5 V nominal, 1 A (short-circuit protected), and 12 V output at 1 A |
External DC Supply
DC-IN Voltage |
15 VDC to 18 VDC (min. 50 W) |
Connector |
Circular, 2.5 mm pin diameter, center-positive |
Battery
Type |
Lithium-ion battery |
Capacity |
93 Wh |
Number of Batteries |
2 |
Life |
5 hours using 2 batteries (hot-swap capable) |
PA/UT Configuration
Bit Depth |
16 bits |
Maximum PRF |
12 kHz |
Frequency
Effective Digitizing Frequency |
Up to 100 MHz |
Display
Refresh Rate |
A-scan: 60 Hz, S-scan: 20 Hz to 30 Hz |
Envelope (echo-dynamic mode) |
Yes: Volume-corrected S-scan (30 Hz) |
A-Scan Height |
Up to 800% |
Synchronization
On Internal Clock |
1 Hz to 10 kHz |
External Pace |
Yes |
On Encoder |
On 2 axes: from 1 to 65,536 steps |
Data Specifications
Processing
Maximum Number of A-Scan Data Points |
Up to 16,384 |
Real-Time Averaging |
PA: 2, 4, 8, 16 UT: 2, 4, 8, 16, 32, 64 |
Rectification |
RF, full wave, half wave+, half wave- |
Filtering |
PA channel: 3 low-pass, 6 band-pass, and 4 high-pass filters UT channel: 8 low-pass, 6 band-pass, and 4 high-pass filters (3 low-pass filters when configured in TOFD) |
Video Filtering |
Smoothing (adjusted to the probe frequency range) |
Programmable TCG
Number of Points |
32: One TCG (time-corrected gain) curve per focal law |
Range |
PA (standard): 40 dB per step of 0.1 dB PA (extended): 65 dB per step of 0.1 dB UT: 100 dB per step of 0.1 dB |
Maximum Slope |
PA (standard): 40 dB/10 ns PA (extended): 0.1 dB/10 ns UT: 40 dB/10 ns |
|
Acoustic Specifications
PulserPA ChannelUT Channels
Voltage |
40 V, 80 V, and 115 V |
85 V, 155 V, and 295 V |
Pulse Width |
Adjustable from 30 ns to 500 ns, resolution of 2.5 ns |
Adjustable from 30 ns to 1,000 ns, resolution of 2.5 ns |
Fall Time |
<, 10 ns |
<, 10 ns |
Pulse Shape |
Negative square pulse |
Negative square pulse |
Output Impedance |
28 in pulse-echo 24 in pitch-catch |
<, 30 |
ReceiverPA ChannelUT Channels
Gain Range |
0 dB to 80 dB maximum input signal, 550 mVp-p (full-screen height) |
0 dB to 120 dB maximum input signal, 34.5 Vp-p (full-screen height) |
Input Impedance |
57 ±, 10% at 9 MHz in pulse-echo 100 ±, 10% at 9 MHz in pitch-catch |
50 in pulse-echo mode 50 in pulse-receive mode |
System Bandwith |
0.5 MHz to 18 MHz |
0.25 MHz to 28 MHz |
Beam FormationPA ChannelUT Channels
Scan Type |
Single, linear, sectorial, compound, and TFM |
– |
Maximum Aperture |
OMNIX3-PATFM1664PR = 16 elements OMNIX3-PATFM16128PR = 16 elements OMNIX3-PATFM32128PR = 32 elements |
– |
Number of Receiving Elements |
OMNIX3-PATFM1664PR = 64 receiving elements OMNIX3-PATFM16128PR = 128 receiving elements OMNIX3-PATFM32128PR = 128 receiving elements |
– |
Number of Focal Laws |
Up to 1024 |
– |
Delay Range Transmission |
0 s to 10 s in 2.5 ns increments |
– |
Delay Range Reception |
0 s to 6.4 s in 2.5 ns increments |
– |
|
TFM/FMC
Supported Modes |
Pulse-echo: L-L, T-T, and TT-TT Self-tandem: LL-L, TT-T, TT-L, TL-T, LT-T, and TL-L |
Number of Groups |
Up to 4 simultaneous TFM groups |
Maximum Aperture |
OMNIX3-PATFM1664PR = 32 elements extended aperture OMNIX3-PATFM166128PR = 32 elements extended aperture OMNIX3-PATFM32128PR = 64 elements extended aperture |
Image Resolution |
Up to 1024 1024 (1 M points) for each TFM group |
Live TFM Envelope |
Yes |
Operating Environment
Ingress Protection Rating |
IP65 certified (completely protected against dust and water jets from all directions (6.3 mm nozzle)) |
Shockproof Rating |
Drop tested according to MIL-STD-810G |
Intended Use |
Indoor and outdoor use |
Altitude |
Up to 2,000 m (6,561 ft) |
Operating Temperature |
0 C to 45 C (32 F to 113 F) |
Storage Temperature |
20 C to 60 C (4 F to 140 F) (with battery inside) 20 C to 70 C (4 F to 158 F) (with no battery inside)
|
|
Package Includes:
– Olympus OmniScan SX 16:64 PA Ultrasonic Flaw Detector (BC# -L/T*)
– Battery Pack
– AC Adapter/Charger
– Power Cord- User Manual On Cd-ROM
– Carrying Case (G)
Reviews